Description
Manufacturer: Novah Enterprises
NovahDent Bowie & Dick test packs fulfills the requirements of EN285, EN ISO 17665, EN ISO 1140-4 and HTM 01-05 for detection of air leaks, steam penetration, inadequate air removal and steam penetration for vacuum assisted steam sterilizers. Single use only. Non toxic. 134 ℃ to 137 ℃ in 3.5 minutes.
- Clear & accurate colour changes (Blue to Black)
- Easy to interpret and record.
EN ISO 17665; 12.1.6 Process control and monitoring: “If the sterilization process relies on the removal of air from the sterilizer chamber in order to achieve rapid and even penetration of steam into the sterilizer load, a steam penetration test shall be carried out each day before the sterilizer is used.”
EN ISO 17665; A.6.2 Routine monitoring and testing: “An alternative test pack may be used providing the challenge to the sterilization process has been demonstrated to be equal to or greater than the challenge from the standard test pack described in A.3.”
Size: 125 x 115 x 20 mm (Actual dimensions of Test Pack)
Packing: Each box contains 20 test packs.